Ht-instruments HT9022 Manuale Utente Pagina 15

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 56
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 14
HT9022
EN - 13
5.2.3. Log
By selecting “Log”, the screen to the side will appear.
Press keys F2, F3 (T, S) to change the duration of the
integration period. It will take the following values:
1, 5, 10, 30, 60, 120, 300, 600 or 900 seconds.
Press key F4 (OK) to save the changes made and go back to
the previous screen (see par.5.2).
OK
Int.Period:
001 s
19/01-17:00:00
5.2.4. InRush
By selecting “InRush”, the screen to the side will appear.
Press key F1 (Sel) to move the cursor and keys F2, F3 (T, S)
to change the selected item.
Press key F4 (OK) to save the changes made and go back to
the previous screen (see par.5.2).
Thresh: threshold value beyond which the inrush current
event is detected and recorded by the instrument. The
current value may be set between 5A and 900A in steps of
1A.
Sel
OK
Thres: 080
Window: 1/1
Mode: Fix
19/01-17:00:00
Window: the value of the inrush current measuring window. The following values are
available:
¾ 1/1: sampling occurs every half-period;
¾ 1/2: sampling occurs one every two half-periods;
¾ 1/4: sampling occurs one every four half-periods;
Mode: the inrush current measuring mode. Following modes are available:
¾ Fix;
¾ Var.
For details about the measurement of Inrush Currents see par. 5.8.2.
5.2.5. Continuity
By selecting “Continuity”, the screen to the side will appear.
Press keys F2, F3 (T, S) to change the setting of the resistance
limit value below which the buzzer will sound. It may be set
between 1 and 150 in steps of 1.
Press key F4 (OK) to save the changes made and go back to
the previous screen (see par.5.2).
Sel
OK
Res Lim:
010
19/01-17:00:00
Vedere la pagina 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 55 56

Commenti su questo manuale

Nessun commento