MACROTESTG3 - COMBIG3
EN - 19
6.2.1. Anomalous situations
1. If the instrument measures a resistance higher than
the set limit value but for which it is not able to generate
the rated voltage, the screen to the side is displayed.
The symbol is shown on the display and the values
of the real test voltage are indicated in red.
2. If the instrument detects a voltage value higher than
10V at its terminals, it does not carry out the test, gives
out a long sound and the screen reported here to the
side is displayed.